This paper proposed a wireless power harvesting system with micro-electro-mechanical-systems (MEMS) fabrication for noninvasive intraocular pressure (IOP) measurement on soft contact lens substructure. The power harvesting IC consists of a loop antenna, an impedance matching network and a rectifier. The proposed IC has been designed and fabricated by CMOS 0.18um process that operates at the ISM band of 5.8 GHz. The antenna and the power harvesting IC would be bonded together by using flip chip bonding technologies without extra wire interference. The circuit utilized an impedance transformation circuit to boost the input RF signal that improves the circuit performance. The proposed design achieves an RF-to-DC conversion efficiency of 35% at 5.8 GHz.

Original languageEnglish
Title of host publication2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2013
Pages3447-3450
Number of pages4
DOIs
StatePublished - 31 Oct 2013
Event2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2013 - Osaka, Japan
Duration: 3 Jul 20137 Jul 2013

Publication series

NameProceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS
ISSN (Print)1557-170X

Conference

Conference2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2013
CountryJapan
CityOsaka
Period3/07/137/07/13

Fingerprint Dive into the research topics of 'The rectenna design on contact lens for wireless powering of the active intraocular pressure monitoring system'. Together they form a unique fingerprint.

  • Cite this

    Cheng, H. W., Jeng, B. M., Chen, C. Y., Huang, H. Y., Chiou, J-C., & Luo, C. H. (2013). The rectenna design on contact lens for wireless powering of the active intraocular pressure monitoring system. In 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2013 (pp. 3447-3450). [6610283] (Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS). https://doi.org/10.1109/EMBC.2013.6610283