Calibration strategy of the JUNO experiment

JUNO Collaboration

Research output: Contribution to journalArticlepeer-review

Abstract

We present the calibration strategy for the 20 kton liquid scintillator central detector of the Jiangmen Underground Neutrino Observatory (JUNO). By utilizing a comprehensive multiple-source and multiple-positional calibration program, in combination with a novel dual calorimetry technique exploiting two independent photosensors and readout systems, we demonstrate that the JUNO central detector can achieve a better than 1% energy linearity and a 3% effective energy resolution, required by the neutrino mass ordering determination.

Original languageAmerican English
Article number4
Number of pages33
JournalJournal of High Energy Physics
Issue number3
DOIs
StatePublished - 1 Mar 2021

Keywords

  • Neutrino Detectors and Telescopes (experiments)